Please ensure you use these settings to properly integrate the Brecknell 6700U Series scale to your APOS system
- Set the protocol as SCP02
- Databits =8
- Parity = none
- Stop bit = 1
- Baudrate : 9600
Below are the steps to navigating to the correct setttings on the brecknell 6700U Series Scale (pay close attention to the bolded steps)
- Power on the indicator by pressing and holding the [ON/OFF] key. The display will go through a display test then go into regular weigh mode.
- Press the [HOLD] + [ON/OFF] key for 3 seconds. The indicator shows confg to indicate that you are in Setup Menu mode.
- Press the [UNIT] key until the indicator display shows user.
- Press the [TARE] key and the indicator display shows reset.
- Press the [UNIT] key until you see the com port you are using for your software; com1 (USB port) or com2 (serial port)
- Press the [TARE] key to select the desired com port.
- The indicator's display shows bud.rt. Press the [TARE] key go into the baud rate setting and confirm it is approprioate baud rate for your software. To toggle to another baud rate setting, press the [UNIT] key until the one is displayed. (i.e. 1200, 2400, 4800, 9600, 19200, 38400)
- Press the [TARE] key and the display shows bud.rt. Press the [UNIT] key and the display shows bt.fmt. Press the [TARE] key to go into the setting and confirm it is the approprioate bit format for your software. To select another bit format setting, press the [UNIT] key until the desired selection is displayed. (i.e. 8n1, 7E1, 7O1)
- Press the [TARE] key and the display shows bt.fmt. Press the [UNIT] key until the display shows lyout.
- Press the [TARE] key to go into your selection and press the [UNIT] key until the display shows correct protocol needed for your software. *Most common as follows scp02 = ECR protocol
- Press the [TARE] key and lyout will be displayed.
- Press the [ZERO] key to display the com port that you were changing.
- Press the [UNIT] key until the display shows other.
- Press the [TARE] key until the display shows nld.rg.
- Press the [UNIT] key until legacy is displayed.
- Press the [TARE] key and a number will be displayed.
- Press the [UNIT] key until a 2 is displayed.
- Press the [TARE] key and legacy will be displayed.
- Press and release the [ZERO/ON/OFF] key until you are all the way out of the Setup Menu mode and into the Normal Weigh mode.
- Test scale with software. If scale fails communication, contact Brecknell Tech Support at 800-242-2807 for further testing.